 With the compact size of Solartron gauging probes and the flexibility of Solartron Metrology’s Orbit® 3 Digital Measurement System. For applications where a contact gauging sensor is unsuitable, this is a cost efficient non-contact measurement solution that includes an adjustable light source, standoff distances and a small spot size.
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 Compact, 8 mm diameter Transducer Gauge Head
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 1.5 mm or 5 mm measuring range
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 8 mm or 24 mm standoff distance
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 Repeatable measurements to ±1 micron
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 30 micron diameter spot size
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 Excellent for measurement on reflective surfaces
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 Adjustment via LCD touch screen
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 Plugs into Orbit® 3, network for up to 150 sensors
http://www.solartronmetrology.com/image.asbx?AttributeFileId=4ab00fc2-d8b6-49aa-9578-2af8f84bffd2 Clear Material Thickness Measurement
 

Confocal Head Type

Measurement Performance

C8H/8/1

C8H/24/5

Calibrated Range mm

1.5

5

Standoff (to start of measuring range) mm

8

24

Linearity (full range) (1) ±%FSO

0.4

0.2

µm

5

10

Linearity (Limited range) (2) ±%FSO

0.2

0.1

µm

2.5

5

Resolution µm

1

1

Repeatability (1) µm

2

2

Spot Diameter µm

30

30

Operating Angle ±°

5

3

Temperature Coefficient µm/°C

2

2

Environmental

 

Operating Temperature °C

20 ±5

Storage Temperature °C

10 to 50

Operating Humidity

Do not use controller in wet conditions

Shock and Vibration

Do not expose controller to shock and vibration

Electrical Interface

 

Power requirements

24V DC from power pack provided

Signal Interfaces

Orbit® 3

Reading Rate R/s

4000

Bandwidth

80

Functional

 

Light Output Level

8 settings to accommodate different levels of reflective surfaces (set from touch screen or Orbit®)

Exposure Settings

5 ms to 100 ms to accommodate different levels of reflective surfaces (set from touch screen or Orbit®)

Averaging

1 to 256 set higher to improve signal to noise ratio

Zero or Absolute

Set from touch screen

Menu

Touch Screen for setting up

Indications

Displayed Measurement and Signal Strength